U-series isotope analysis




top

Analytical issues and challenges


  • traditional α-counting yields a precision of ∼4-10%
  • Th and Pa have thermal ionisation efficiency <0.01% (PIMS)
  • Th and Pa are highly insoluble
  • Ra has a thermal ionisation efficiency ∼5-14% (TIMS)
  • Ra and Pa only present at fg/g levels
  • 234U/238U ratios ∼10-5, 230Th/232Th ratios ∼10-6
  • 238U and 232Th "tail" onto 234U and 230Th, respectively

Spikes

  • Mixed 236U-229Th spike for U-Th analysis (Oak Ridge)
  • 228Ra spike milked from equilibrium 232Th solution (NIST3159)
  • 233Pa spike milked from 237Np or derived by irradiation of 232Th
  • 209Po spike available from NIST
  • Calibration against SRM solution standards
  • Cross-check with analysis of secular equilibrium material (TML)
  • Th and Pa spikes need ∼0.01M HF to keep in solution
Decay chains of the U-series isotopes.

Decay chains of the U-series isotopes (activity = λN). Click to enlarge


top

Analytical facilities

Laboratories


  • Newly constructed clean lab facilities with four self-contained laboratories, including two dedicated solely for U-series work.
    The clean lab area includes 7 laminar flow workstations and two separate balance rooms
Laboratories

U-series clean lab facilities

  • Nu Plasma high resolution multi-collector ICPMS, equipped with an RPQ filter (Installed Nov. 2003)
  • Sample in solution of 2% HNO3, sprayed into the plasma using a DSN 100
Nu Plasma MC-ICP-MS

Nu Plasma MC-ICP-MS

  • Finnigan Triton TIMS (Installed March 2005) Equipped with RPQ filter, SEM and oil-free scroll pump.
  • Ra samples loaded onto outgassed, zone-refined Re filaments using a Ta-HF-H3PO4 activator solution
Triton

Finnigan Triton TIMS


top

U isotope analysis by MC-ICP-MS


  • U analysis using CRM145 or CRM960 standard
  • Comparison of measured 238U/235U ratio to the true ratio (137.88) yields mass bias fractionation.
  • Comparison of mass fractionation-corrected 234U/238U to the true ratio (5.286x10-5) yields ion counter gain
  • Intersperse samples with CRM145 analyses

  • Cycle 1

  • 234U on Ion Counter 0 (IC0)
  • 235U on Faraday cup L2
  • 238U on Faraday cup H1

  • Cycle 2

  • 236U on IC0
  • 238U on Faraday cup L1
Standard measurements of TML

234U/238U standard measurements of TML. True value= 1. Click to enlarge.


top

Th isotope analysis by MC-ICP-MS


  • A solution of 20ppb Th yields a beam of and ∼1V of 232Th, 100s of cps of 230Th
  • 230Th is collected on an ion counter preceded by a RPQ deceleration lens. Typical abundance sensitivity is ∼450ppb at 1 a.m.u. for a transmission of 95-98% (190ppb at 2 a.m.u. but transmission not determined)

  • Th analysis using CRM145 or CRM960 and Th 'U' standards
  • Comparison of measured 238U/235U ratio to the true ratio (137.88) yields mass bias fractionation.
  • Comparison of mass fractionation-corrected 234U/238U to the true ratio (5.286x10-5) yields ion counter gain
  • Intersperse samples with Th 'U' analyses to get βTh

Standard measurements of TML

TML-230Th/232Th bracketed with Th'U'. True value= 5.80x10-6. Click to enlarge.

    Cycle 1

  • 230Th on IC0
  • 232Th on Faraday cup L1
  • Cycle 2

  • 229Th on IC0
  • 232Th on axial Faraday cup (Ax)
Standard measurements of TML

TML-230Th/238U bracketed with Th'U'. True value= 1. Click to enlarge


top

Ra isotope analysis by TIMS

  • Slow heating to 1200°C then use 138Ba to focus
  • Scan masses 224-230 to check for Ra and hydrocarbons
  • Build Ra beam (100-1000 cps @ 1250-1350°C)
  • Dynamic collection on SEM ∼80 ratios with 8 second integration times (precision <0.5%)
TML for<sup> 226</sup>Ra/<sup>230</sup>Th

TML standard measurements for 226Ra/230Th. True value= 1. Click to enlarge.


top

210Pb isotope analysis by α-counting

  • 210Pb concentrations are determined from 210Po on separate 1-2 g sample dissolutions which are spiked with a 209Po tracer calibrated against a value of 7.95 dpm/g for TML
  • Samples are loaded onto a 1 ml anionic column in 1N HCl containing H3BO3 where most major and trace elements were washed off in 1N and 0.5N HCl after which Po is eluted in warm 7.5N HNO3
  • Po is autoplated onto a Ag disc above a magnetic stirrer for 6-8 hours in 150 ml of warm 0.5N HCl which contained ascorbic acid. 210Po and 209Po were counted for ∼1 week to achieve a precision of <4%
α-counter

Ortec Octete Plus α-counter.